CALIBRATING THE INSTRUMENT DO 9704

Apply a voltage of 465.5 mVdc between pin 6 (GND) and pin 1 (OUT) of the DIN 8-pole connector at input A.

Instrument

 

 

DO 9704

 

 

 

7

6

B

A

 

Vdc=465.5

 

3

8

1

 

 

 

 

54

2

Select step 11 with the PROG key; press the ENTER key, the display shows

Using the sand tkeys, make the value shown on the display coincide with 196.0. Confirm by pressing the ENTER key; P11 appears on the display; press PROG: P0 appears; press ENTER: this quits the instrument calibration function.