FLOW RATE MEASUREMENTS

A selection resistance sets the conversion factor of the connected turbine.

The instrument automatically recognises 5 fixed flow rate values plus a sixth which may be configu- red by the user.

FLOW RATE PROBES

 

Measuring range

Selection

Output

Indication

 

 

resistance k

frequency PPL

lt/min.

 

 

 

 

 

LT 5

5 l/min.

90.9

5

5

(LT10)*

 

95.3

5

20

LT20

20 l/min.

 

 

 

(LT50)*

 

100

0.5

125

LT125

125 l/min.

 

 

 

(LT250)*

 

105

0.1

400

(LT300)*

 

 

 

 

LT400

400 l/min.

 

 

 

LT500

500 l/min.

110

0.04

500

LT750

750 l/min.

115

0.04

750

 

 

 

 

 

LT configurable

2000 l/max

121

configurable

PPL

*The instrument can read a reduced measuring range, it has the same selection resistance.

The connection cable of the configurable turbine will end with a DIN 8-pole connector; connect a 121 kresistance between pin 3 and pin 7.

1

2

3

4

5

6

7

8

6

7

13

8

45

2

N.C.

N.C.

Select.

 

N.C.

N.C.

 

 

 

 

 

 

 

 

 

R.

 

 

 

*

Active probe for measuring flow rate with a turbine 2 to 2000 litres/minute.

*A coil which generates impulses with a minimum of 100 mV p-p is connected to pins 4 and 6.