Emerson 00809-0100-4514, 1154 Installation, Procedures, Mechanical, Rosemount, Reference Manual

Models: 1154 00809-0100-4514

1 60
Download 60 pages 13.73 Kb
Page 18
Image 18
INSTALLATION

Rosemount 1154

Reference Manual

00809-0100-4514, Rev BA January 2008

INSTALLATION

Installation consists of mounting the transmitter and conduit and making

PROCEDURES

electrical connections. Procedures follow for each operation.

Mechanical

Transmitter

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Be careful not to break the neck seal between the sensor module and the

 

electronics housing.

 

 

 

The threaded interface between the sensor module and the electronics

 

housing is hermetically sealed before shipment. The integrity of this seal is

 

necessary for the safe operation of the transmitter during accident conditions.

 

If the seal is broken, reseal it according to “Connecting the Electrical Housing

 

to the Sensor Module” on page 5-7.

1.Mount the bracket to a panel or other flat surface as shown in Figure 2-4.Use four 3/8-in. diameter bolts (not supplied with unit). SAE grade 2 bolts were used during qualification testing. Torque each bolt to 19 ft-lb (26 N-m).

2.Attach the transmitter to the mounting bracket, as shown in Figure 2-4.Use four 7/16-20 ￿ 3/4 bolts with washers (supplied with unit). Torque each bolt to 21 ft-lb (29 N-m).

Figure 2-4. Typical Transmitter Mounting Bracket Configuration.

2.81 (71.4)

 

1.2

 

(30)

 

4.93

 

(125)

2.81

3/8-in.

(71.4)

Bolts (4)

5

(Customer

(127)Supplied)

2.3Center of Gravity

(58)(Includes Bracket)

1.8

(45.7)

2.75

(69.9)

10 (254)

Minimum Clearance

2.3

(58)

Center of Gravity (Includes Bracket)

PANEL MOUNTING HOLE

MOUNTING BRACKET FOR PANEL

ACCEPTABLE ALTERNATE

PATTERN

MOUNT SHOWN IN TYPICAL

MOUNTING

(BACK SIDE)

MOUNTING CONFIGURATION

 

NOTE

Dimensions are nominal in inches (millimeters).

NOTES

1.Orientation with respect to gravity is not critical.

2.Units can alternately be mounted with process connection adjacent to bracket.

2-6