Reference Manual

00809-0100-4514, Rev BA January 2008

Rosemount 1154

Section 5 Maintenance and

Troubleshooting

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-1

Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-2

Test Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-2

Board Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-2

Sensing Module Checkout . . . . . . . . . . . . . . . . . . . . . . . . . page 5-3

Disassembly Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-4

Reassembly Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-6

Post Assembly Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-9

 

 

 

 

 

 

 

 

 

 

 

 

 

Use only the procedures and new parts specifically referenced in this manual to

 

ensure specification performance and certification compliance. Unauthorized

 

procedures or parts can render the instrument dangerous to life, limb, or property.

 

 

 

 

OVERVIEW

This section outlines a technique for checking out the components, a method

 

for disassembly and reassembly, and a troubleshooting guide.

 

 

 

 

 

NOTE

 

Maintenance of traceability of any replacement parts is the responsibility of

 

the user (see “Important Notice” on page 6-11and Important Notice at the

 

beginning of this manual preceding the Table of Contents).

 

 

 

 

 

The Rosemount 1154 has no moving parts and requires a minimum of

 

scheduled maintenance. Calibration procedures for range adjustment are

 

outlined in Section 3: Calibration. A calibration check should be conducted

 

after inadvertent exposure to overpressure, unless your plant considers this

 

factor separately in the plant error analysis.

 

Test terminals are available for in-process checks. For further checks, the

 

transmitter can be divided into three active physical components: the sensing

 

module, the amplifier board, and the calibration board.

 

An exploded view of the transmitter is provided in Figure 5-2 on page 5-5.In

 

the following procedures, numbers in parentheses refer to item numbers in the

 

exploded view.