IMPORTANT NOTICE -- ERRATA

 

 

 

Model 1154 Product Manual 00809-0100-4514 Rev BA (January 2008)

 

 

 

 

 

 

No.

Affected

 

Description of Change

Effect.

 

Pages

 

 

Date

1

6-6

Process Flange CF3M(Cast version of 316L SST)

10/21/09

 

 

Drain/Vent Valves 316LSST

 

 

 

Process Connections – 3/8-inch Swagelok compression fitting, 316L SST (1/4-18

 

 

 

NPT optional)

 

 

2

3-5

Change the last paragraph on the page to read as follows:

4/13/12

 

 

”Damping electronics are available as an option. Transmitters with standard

 

 

 

electronics can be retrofitted with the adjustable damping feature by changing out

 

 

 

both the amplifier board and the calibration board. Please reference Table 6-2

 

 

 

(Rosemount 1154DP, 1154HP, and 1154GP Spare Parts) for the applicable part

 

 

 

numbers.”

 

 

 

 

 

 

3

6-9

Table 6-2,”Rosemount 1154DP, 1154HP, and 1154GP Spare Parts”, in all locations

4/13/12

 

 

the following part numbers are updated:

 

 

 

Amplifier Circuit Board, Output Code R:

 

 

 

”01154-0001-0005” is replaced by ”01154-0153-0001”

 

 

 

Amplifier Circuit Board with Damping, Output Code R:

 

 

 

”01154-0021-0004” is replaced by ”01154-0156-0001”

 

 

 

Amplifier Circuit Board, Output Code R, N0026:

 

 

 

”01154-0001-0006” is replace by ”01154-0153-0002”

 

 

 

Sensor Module, 316 SST (5):

 

 

 

0-25/150 inH2O: ”01154-0300-0242” is replaced by ”01154-5300-0242”

 

 

 

 

”01154-0300-0342” is replaced by ”01154-5300-0342”

 

 

 

 

”01154-0300-0142” is replaced by ”01154-5300-0142”

 

 

 

0-125/750 inH2O: ”01154-0300-0252” is replaced by ”01154-5300-0252”

 

 

 

 

”01154-0300-0352” is replaced by ”01154-5300-0352”

 

 

 

 

”01154-0300-0152” is replaced by ”01154-5300-0152”

 

 

 

0-17/100 psi:

”01154-0300-0262” is replaced by ”01154-5300-0262”

 

 

 

 

”01154-0300-0362” is replaced by ”01154-5300-0362”

 

 

 

 

”01154-0300-0162” is replaced by ”01154-5300-0162”

 

 

 

0-50/300 psi:

”01154-0300-0272” is replaced by ”01154-5300-0272”

 

 

 

 

”01154-0300-0372” is replaced by ”01154-5300-0372”

 

 

 

 

”01154-0300-0172” is replaced by ”01154-5300-0172”

 

 

 

0-170/1,000 psi:

”01154-0300-0282” is replaced by ”01154-5300-0282”

 

 

 

 

”01154-0300-0182” is replaced by ”01154-5300-0182”

 

 

 

0-500/3,000 psi:

”01154-0300-0192” is replaced by ”01154-5300-0192”

 

0-1,000/4,000 psi: ”01154-0300-0102” is replaced by ”01154-5300-0102”

4 6-10

Table 6-2,” Rosemount 1154DP, 1154HP, and 1154GP Spare Parts”, the table

4/13/12

 

footnotes are updated to add note (5) which will read as follows:

 

 

(5) IMPORTANT NOTICE: To maintain a transmitter’s qualified configuration,

 

 

when purchasing or installing a new Sensor Module, Rosemount Qualification

 

 

report D2011019 must be carefully reviewed to verify that the Sensor Module

 

 

to be installed and the associated Amplifier Circuit Board in a given

 

 

transmitter is a qualified configuration. As detailed in the referenced report, not

 

 

all Sensor Module part numbers are considered qualified in conjunction with certain

 

 

Amplifier Circuit Boards.

 

Page 2
Image 2
Emerson 00809-0100-4514, 1154 manual Important Notice -- Errata