Reference Manual
Rosemount 1154
Figure 3-2. Zero and Span Adjustment Screws.
Zero |
Span |
Example (for Range Code 4)
Initial transmitter calibration: 25 to 125 inH2O
(100 inH2O span with zero suppressed 25 inH2O)
Desired transmitter calibration:
1.Adjust the zero to eliminate any existing zero elevation or suppression. With 0 inH2O pressure applied to the transmitter, turn the zero adjustment until the output reads 4 mA. The unit is now calibrated for 0 to 100 inH2O.
2.Adjust the span to the desired new span. To reduce the span, turn the span screw until the output, with 0 inH2O pressure input, equals:
4 mA × | Existing Span | = 4 mA × | 100 inH2O | = 8 mA | |
|
|
|
3.Adjust the zero screw to bring the output, with 0 inH2O input, back to 4 mA. The transmitter calibration should now be very close to 0 to 50 inH2O.
4.Check
For example, if, after completing step 3, the transmitter output reads
19.900 mA at 50 inH2O, turn the span potentiometer until the output (at 50 inH2O) reads 20.025 mA.
19.900 + (20.000 – 19.900) 1.25 =
19.900 + 0.125 = 20.025
Since the span adjustment affects zero 1/5 as much as the span, the 0.125 mA increase in span causes a 0.025 mA increase in zero.
Therefore, turn the zero adjustment (at 50 inH2O) until the output reads 20.000 mA. The unit should now be calibrated for 0 to 50 inH2O.
5.Zero Elevation/Suppression. Elevate zero. Turn the screw until the
output reads 4 mA with