Rosemount 1154
Reference Manual
Static Pressure and Overpressure Limits
Rosemount 1154DP:
0.5psia to 2,000 psig (3.4 kPa abs to 13.8 MPa) maximum rated static pressure for operation within specifications; overpressure limit is 2,000 psig (13.8 MPa) on either side without damage to
the transmitter.
Rosemount 1154HP:
0.5psia to 3,000 psig (3.4 kPa abs to 20.7 MPa) maximum rated static pressure for operation within specifications; overpressure limit is 3,000 psig (20.7 MPa) on either side without damage to
the transmitter.
Overpressure Limits
Rosemount 1154GP:
Operates within specifications from 0.5 psia
(3.4 kPa abs) to upper range limit. Overpressure limits without damage to the transmitter:
Range Code | Overpressure Limit |
|
|
2,000 psig (13.8 MPa) | |
9 | 4,500 psig (31.0 MPa) |
|
|
0 | 6,000 psig (41.34 MPa) |
|
|
PHYSICAL SPECIFICATIONS
Materials of Construction
Isolating Diaphragms:
316L SST
Drain/Vent Valves:
316 SST
Process Flanges:
Process
316L SST
Electronics Housing
Ethylene propylene
Fill Fluid:
Silicone oil
Flange Bolts and Nuts:
Plated alloy steel, as specified in ASTM A540
Electronics Housing:
316 SST
Mounting Bracket:
316L SST
Mounting Bolts (Bracket to transmitter): SAE J429 carbon steel, Grade 2 or Grade 5