Static Pressure Zero Effect

Rosemount 1154DP:

Per 1,000 psi (6.89 MPa):

Range Code

Static Pressure Zero Effect

 

 

4, 5

±0.2% of upper range limit

6–8

±0.5% of upper range limit

 

 

FUNCTIONAL SPECIFICATIONS

Rosemount 1154HP:

Per 1,000 psi (6.89 MPa):

Range Code

Static Pressure Zero Effect

 

 

All Ranges

±0.66% of upper range limit

 

 

Static Pressure Span Effect

Effect is systematic and can be calibrated out for a particular pressure before installation. Correction uncertainty is ±0.5% of input

reading/1,000 psi (6.89 MPa).

Power Supply Effect

Less than 0.005% of output span/volt

Load Effect

No load effect other than the change in voltage supplied to the transmitter

Mounting Position Effect

No span effect; zero shift of up to 1.5 inH2O (372 Pa) which can be calibrated out

Response Time

Fixed time constant (63%) at 100 °F (37.8 °C) as follows:

Range Code

Response Time

 

 

4

0.5 seconds or less

all others

0.2 seconds or less

 

 

Adjustable damping is available through a special

N option.

Service

Liquid, gas, or vapor

Output

4–20 mA dc

Power Supply

Design limits are as shown in Figure 2-3 on page 2-5. See qualification report D8400102 for additional detail.