P =

Rosemount 1154

Reference Manual

00809-0100-4514, Rev BA January 2008

K1

is a constant.

C1

is the capacitance between the high pressure side and the sensing

 

diaphragm.

C2 is the capacitance between the low pressure side and the sensing diaphragm.

fVp p

Iref

= --------------------

 

C1 + C2

Where:

Iref is the current source.

Vp-pis the peak-to-peak oscillation voltage.

fis the oscillation frequency.

Idiff = fVp p(C2C1)

Where:

Idiff is the difference in current between C1 and C2.

Therefore:

P =

Constant × I

 

= I

C2 C1

 

--------------------

 

 

diff

 

refC2 + C1

Figure 4-1. The δ-Cell.

Leadwires

Isolating

Diaphragm

Capacitor

Plates

Sensing

Diaphragm

Rigid

Insulation

Silicone Oil

Welded Seals

4-2