Emerson 1154 Rosemount, Reference Manual, ±0.25% of upper range limit, ±1.0% of upper range limit

Models: 1154 00809-0100-4514

1 60
Download 60 pages 13.73 Kb
Page 47
Image 47
±0.25% of upper range limit

Reference Manual

00809-0100-4514, Rev BA January 2008

Rosemount 1154

Drift

±0.2% of upper range limit for 30 months

Range Code 0: ±(0.3% of upper range limit)

Temperature Effect

Range Codes 4–9:

±(0.75% upper range limit +0.5% span) per 100 °F (55.6 °C) ambient temperature change

Range Code 0:

±(1.13% upper range limit +0.5% span) per 100 °F (55.6 °C) ambient temperature change

Overpressure Effect

Rosemount 1154DP:

Maximum zero shift after 2,000 psi (13.8 MPa) overpressure:

Range Code

Overpressure Effect

 

 

4

±0.25% of upper range limit

5

±1.0% of upper range limit

 

 

6, 7

±3.0% of upper range limit

8

±6.0% of upper range limit

 

 

Rosemount 1154HP:

Maximum zero shift after 3,000 psi (20.68 MPa) overpressure:

Range Code

Overpressure Effect

 

 

4

±1.0% of upper range limit

5

±2.0% of upper range limit

 

 

6, 7

±5.0% of upper range limit

 

 

Rosemount 1154GP:

Maximum zero shift after 2,000 psi (13.8 MPa) overpressure:

Range Code

Overpressure Effect

 

 

4

±0.25% of upper range limit

5–8

±1.0% of upper range limit

 

 

After 4,500 psi (31.0 MPa) overpressure:

Range Code

Overpressure Effect

 

 

9

±0.5% of upper range limit

 

 

After 6,000 psi (41.37 MPa) overpressure:

Range Code

Overpressure Effect

 

 

0

±0.25% of upper range limit

 

 

6-3

Page 47
Image 47
Emerson 1154, 00809-0100-4514 manual Rosemount, Reference Manual, ±0.25% of upper range limit, ±1.0% of upper range limit