Emerson Process Management CI-9110 Detailed Calibration For Analog Model, Feedback Zero & Span

Models: CI-9110 9110-00A Series 9110

1 74
Download 74 pages 60.74 Kb
Page 46
Image 46

5.If above readings are within stated specifications, stop! No calibration is required. Restore instrument to normal operation. Otherwise, proceed to topic 5.7 or 5.8 as required.

5.7 DETAILED CALIBRATION FOR ANALOG MODEL

If calibration errors were found in topic 5.6, the procedures described here for Analog models are performed.

5.7.1 Feedback Zero & Span

Precision voltage of an input test circuit is used to calibrate the zero and span as follows:

1.Adjust input test circuit for 1.000 V output.

2.DMM #1 should read 0.995 V, ±.004 V.

3.If reading of step 2 is out of tolerance, slightly loosen two screws on body of feedback pot (Figure 5-1) and rotate body until DMM #1 reads as specified in step 2. Secure pot screws to lock in setting.

4.Adjust the input test circuit for 5.000 V.

5.DMM #1 should read 4.995 V, ±.004 V

6.If the reading of step 5 is out of tolerance, adjust potentiometer R35 on CPU to correct.

7.Recheck both points and readjust zero and span if required.

5.7.2 Current Output Span & Load

The current output circuitry is provided with two calibration adjustments. Potentiometer R50 is used for span adjustment and R49 for load compensation. For these tests, DMM #2 (used as a milliammeter) is connected in series with a 250 ohm load resistor as shown in Figure 5-2. The pushbutton switch is used for load testing.

1.Adjust input test circuit for 5.000 V.

2.DMM #2 should read 20 mA, ±.02 mA. If necessary, adjust pot R50 to correct.

3.Press load pushbutton while observing DMM #2. If current reading increases, adjust R49 CCW. If reading decreases, turn it CW. Repeat procedure until change is less than

±.02 mA.

4.Final reading should be 20 mA, ±.02 mA. If necessary, reset pot R50 and repeat the procedure.

5.7.3 Pressure Output Check

1. Adjust input test circuit for 1.000 V.

Service

5-6

CI-9110

Page 46
Image 46
Emerson Process Management CI-9110 Detailed Calibration For Analog Model, Feedback Zero & Span, Current Output Span & Load