Emerson Process Management 9110-00A, Series 9110 Calibration Test Setup, Calibration Precaution

Models: CI-9110 9110-00A Series 9110

1 74
Download 74 pages 60.74 Kb
Page 44
Image 44

8.Once limits have been set, re-apply dc power and adjust input test circuit (analog or raise/lower) for range value above upper limit. When upper limit is reached, main gear will stop rotating and output pressure will hold constant at value of desired limit.

** Caution **

Do not attempt to turn main gear or driveshaft by hand to reach a limit. The force generated in this manner can easily damage the gear teeth and limit tangs. Always use a dc test input signal to drive the main gear through its range.

9.If limit values are not at desired points, continue to experiment with tangs until correct positions can be found.

10.When adjustments are complete, disconnect test setup and recheck board configuration as required. Replace cover and restore unit to normal operating status.

5.4 CALIBRATION TEST SETUP

The calibration of the Remote Set Regulator will require a laboratory bench setup. The setup of Figure 5-2 is used for Analog models, while that of Figure 5-3 is used for Raise/Lower models.

In Figure 5-2, the Analog model uses a precision 1 - 5 volt test source to provide an ad- justable input signal. In Figure 5-3, the Raise/Lower model uses a switch test circuit con- sisting of two pushbutton switches that apply dc power to either the Raise or Lower inputs. As long as a switch is pressed, the output will be raised or lowered assuming continuous dc input configuration.

From this point on, the test setups of Figures 5-2 and 5-3 are identical. A digital multimeter (DMM #1) is connected across test points TP3 and TP4 to monitor the voltage at the center arm of the feedback potentiometer. The 250 ohm resistor and switch circuit is used to load the current output circuit, while DMM #2,functioning as a milliammeter, measures the load current.

On the pneumatic side, a constant supply pressure is applied to the pressure regulator as- sembly. A precision test gauge monitors the pressure output signal. The pressure regulator assembly of the unit does not have calibration adjustments. All adjustments are electrical.

5.5 CALIBRATION PRECAUTION

During calibration, the stepper motor will be driven to both ends of it operating range. Since this movement is extremely slow, do not be tempted to run the unit up or down range by manually forcing the main gear or driveshaft. Doing so can cause damage to the gear teeth and the mechanical limit tangs. Always use a dc test input signal to reposition the main gear and allow the Remote Set Regulator to balance itself out.

Service

5-4

CI-9110

Page 44
Image 44
Emerson Process Management 9110-00A, Series 9110, CI-9110 instruction manual Calibration Test Setup, Calibration Precaution