Mainframe Commands

MESE<N> (Module Event Status Enable)

Example

OUTPUT XXX;":MESE1 3"

 

 

Query:MESE<N>?

The query returns the current setting. Table 10-8 lists the Module Event Status Enable register bits, bit weights, and what each bit masks for the mainframe.

Returned Format [:MESE<N>] <enable_value><NL>

Example

OUTPUT XXX;":MESE1?"

 

 

Table 10-8

 

HP 16500C Mainframe (Intermodule) Module Event Status Enable Register

 

 

 

 

 

 

Bit Position

Bit Weight

Enables

 

7

128

not used

 

6

84

not used

 

5

32

not used

 

4

16

not used

 

3

8

not used

 

2

4

not used

 

1

2

RNT - Intermodule Run Until Satisfied

 

0

1

MC - Intermodule Measurement Complete

10–17