HP Sander 16500C manual Mesrn Module Event Status Register, Mesrn?

Models: Sander 16500C

1 252
Download 252 pages 12.2 Kb
Page 150
Image 150

Mainframe Commands

MESR<N> (Module Event Status Register)

 

MESR<N> (Module Event Status Register)

Query

:MESR<N>?

 

The MESR query returns the contents of the Module Event Status register.

 

The <N> index specifies the module. For the HP 16500C alone, the <N>

 

index 0 through 5 refers to system and modules 1 through 5 respectively.

 

With an HP 16501A connected, the <N> index 6 through 10 refers to modules

 

6 through 10 respectively.

 

Refer to table 10-9 for information about the Module Event Status Register 0

 

bits and their bit weights.

See Also

MESR in the Programmer’s Guide for a specific module for the

 

interpretation of that module’s Event Status Register.

Returned Format

[:MESR<N>] <enable_value><NL>

<N>

An integer 0 through 1.

<enable_value>

An integer from 0 through 255

 

 

Example

OUTPUT XXX;":MESR1?"

 

 

Table 10-9

 

HP 16500C Mainframe Module Event Status Register (<N>=0)

 

 

 

 

 

 

 

Bit

Bit Weight

Bit Name

Condition

 

7

128

 

not used

 

6

64

 

not used

 

5

32

 

not used

 

4

16

 

not used

 

3

8

 

not used

 

2

4

 

not used

 

1

2

RNT

0 = Intermodule Run until not satisfied

 

 

 

 

 

1

= Intermodule Run until satisfied

 

0

1

MC

0

= Intermodule Measurement not satisfied

 

 

 

 

 

1

= Intermodule Measurement satisfied

10–18

Page 150
Image 150