INTermodule Subsystem

PORTEDGE

Command

<edge_spec>

Example

Query

Returned Format

Example

PORTEDGE

:INTermodule:PORTEDGE <edge_spec>

This command does not obey the truncation rules.

The PORTEDGE command sets the Port In BNC to respond to either a rising edge of falling edge for a trigger from an external source. The threshold level of the input signal is set by the PORTLEV command.

A 1 or ON for rising edge or a 0 or OFF for falling edge.

OUTPUT XXX;":INTERMODULE:PORTEDGE 1"

:INTermodule:PORTEDGE?

The PORTEDGE query returns the current edge setting.

[:INTermodule:PORTEDGE] {10}<NL>

OUTPUT XXX;":INTERMODULE:PORTEDGE?"

13–12

Page 210
Image 210
HP Sander 16500C manual INTermodulePORTEDGE edgespec, INTermodulePORTEDGE?, Portedge query returns the current edge setting