:INTermodule:INSert

 

 

 

 

 

 

{<module>OUT},{GROUP<module>}

 

 

 

The INSert command adds a module or PORT OUT to the Intermodule

 

configuration. The first parameter selects the module or PORT OUT to be

 

added to the intermodule configuration, and the second parameter tells the

 

instrument where the module or PORT OUT will be located. 1 through 5

 

corresponds to the slot location of the modules A through E for the HP

 

16500C alone and 1 through 10 corresponds to slot location of modules A

 

through J when an HP 16501A is connected.

 

 

<module>

An integer, 1 through 5 for HP 16500C alone. 1 through 10 with the HP

 

16501A connected.

 

 

 

 

 

 

 

 

 

 

 

 

Example

OUTPUT XXX;":INTERMODULE:INSERT 1,GROUP"

 

 

 

OUTPUT XXX;":INTERMODULE:INSERT 2,GROUP"

 

 

 

OUTPUT XXX;":INTERMODULE:INSERT 3,2;INSERT OUT,2"

 

The following figure shows the result of the example output commands:

 

 

 

 

Group Run

 

 

 

 

B

 

 

 

 

 

 

 

 

 

 

 

 

A

 

 

 

 

 

 

C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

OUT

13–9