SYSTem Subsystem

SETup

SETup

Command:SYSTem:SETup <block_data>

The :SYSTem:SETup command configures the logic analysis system as defined by the block data sent by the controller. This chapter describes briefly the syntax of the Setup command and query for the mainframe.

Because of the capabilities and importance of the Setup command and query for individual modules, a complete chapter is dedicated to it in each of the module Programmer’s Guides. The dedicated chapter is called "DATA and SETup Commands."

<block_data> <block_length_specifier><section>

<block_length_ #8<length> specifier>

<length> The total length of all sections in byte format (must be represented with 8 digits)

<section> <section_header><section_data>

<section_ 16 bytes, described in the "Section Header Description" section in the "DATA header> and SETup Commands" chapter of the module Programmer’s Guides.

<section_data> Format depends on the type of data

The total length of a section is 16 (for the section header) plus the length of the section data. When calculating the value for <length>, don’t forget to include the length of the section headers.

ExampleOUTPUT XXX USING "#,K";":SYSTEM:SETUP " & <block_data>

11–12