ARMS

PROCEDURE 2

3.Refer to FIGURE 6 to determine the mounting hole in the adjustment plate that will be used to corre- spond to the back angle.

NOTE: Back angles of 105o and 110o will use the same arm adjustment plate mounting holes.

4.Securely tighten the locking pin and washer to the adjustment plate with a locknut.

5.Repeat STEPS 1-4for the opposite side, if neces- sary.

REPLACING/REPOSITIONING THE LOCKING MECHANISM IN THE CANTILEVER ARM (FIGURE 12)

1.Move the cantilever arm up and out of the way.

2.Remove the actuator from the locking mechanism.

CAUTION

The locking mechanism is spring loaded. Place your free hand over the locking mechanism to prevent the parts from springing out of the cantilever arm.

3.Slowly let the locking mechanism and spring slide out of the cantilever arm.

NOTE: Inspect the spring and replace if necessary.

4.Slide the new locking mechanism and spring into the cantilever arm.

5.Position the angled portion of the locking mecha- nism in one (1) of two (2) ways:

Angled Portion Facing Up - Arm will flip UP Angled Portion Facing Down - Arm will flip DOWN

6.Use Loctite 242 and securely tighten the actuator into the locking mechanism.

7.To lock the cantilever arm, push down until there is an audible click.

8.Pull up on the cantilever arm to make sure it is locked in place.

Angled Portion of

Cantilever Arm

Locking Mechanism

Spring

 

Locking

Mechanism

Actuator

(Apply Loctite 242)

FIGURE 12 - REPLACING/REPOSITIONING THE LOCKING MECHANISM IN THE FLIP-UP HALF ARM

ARM PAD DEPTH ADJUSTMENT/ REPLACEMENT - CANTILEVER ARMS (FIGURE 13)

Adjustment

1.Remove the phillips screw from the rear of the arm- rest pad and self-taping screw if FULL LENGTH ARM PADS.

2.Depending on the desired arm pad depth, reposi- tion the cantilever slide tube to one (1) of five (5) positions for DESK LENGTH ARM PADS and into the FIRST adjustment hole for the FULL LENGTH ARM PADS.

3.Reattach the arm pad/cantilever slide tube to the arm tube with existing hardware.

4.Repeat for the opposite side, if necessary.

Replacement

1.Remove the phillips screws from the armrest pad.

2.Replace with NEW armrest pad.

3.Secure with existing hardware.

Desk Length

Arm Pads

Phillips Screw

(Rear of Armrest

Pad) Cantilever Slide Tube

NOTE: This Phillips Screw only needs to be re- moved when Armrest Pad is replaced.

Full Length

Arm Pads

Phillips Screw

(Rear of Armrest

Pad)

Cantilever Arm Self-Tapping Screw NOTE: Only these Adjust- ment Holes can be used.

Cantilever Slide Tube

NOTE: This Phillips Screw only needs to be re- moved when Armrest Pad is replaced.

FIGURE 13 - ARM PAD DEPTH ADJUSTMENT/ REPLACEMENT - CANTILEVER ARMS

A R M S

25

Page 25
Image 25
Invacare 1061057, 96-100 Angled Portion Cantilever Arm, Locking Mechanism Actuator Apply Loctite, Adjustment, Replacement