SECTION 6—OPERATING INSTRUCTIONS

DETAIL “A” DETAIL “B”

Humidifier

To Oxygen Outlet

Bottle Adapter

Connector (Not Shown)

 

Humidifier

 

Bottle

Humidifier

Adapter

Bottle

Humidifier

 

 

Bottle

Patient

Cannula

FIGURE 6.4 Humidifier Attachment

Transportable Oxygen Concentrator Front Panel and LCD Display

NOTE: For this procedure, refer to FIGURE 6.9 on page 30.

The front panel consists of LED indicators (red, yellow and green), control/adjustment buttons and a LCD display.

Indicators and Audible Alarm

There are three indicator LEDs located just above the LCD display ‐ green, yellow and red. The indicators could be on steady or blinking at different rates depending on the specific operating condition, including any warning or alarm.

NOTE: For detailed explanation and descriptions of the warnings and alarms, see

Warnings and Alarms on page 37 or Troubleshooting on page 35.

Green Indicator

The Green indicator is on when the unit is operating normally and providing good (>85%) oxygen purity.

Yellow Indicator

The Yellow indicator is typically on, or blinking, when there is a warning or non‐optimal condition present. The transportable oxygen concentrator could still be operating, but some user action may be required to return to normal operation.

Red Indicator

The Red indicator is typically lit, or flashing, when there is an alarm that requires the immediate attention of the user. Alarms typically result in the unit stopping with no delivery of oxygen.

Part No 1156872

25

Invacare SOLO2

Page 25
Image 25
Invacare TPO100B user manual Indicators and Audible Alarm, Detail a Detail B