SECTION 8—TROUBLESHOOTING GUIDE

LCD DISPLAY:

INDICATORS:

 

 

 

Continuous audible beep every half

 

second.

 

RED indicator is On.

 

GREEN indicator is Off.

 

 

DESCRIPTION:

SOLUTIONS:

 

 

 

1. Make sure that car (boat or motor

 

home, etc.) is running.

External DC Power is too low to allow

2. Make sure DC Power Cable is firmly

continued operation.

connected at both ends.

Excessive current draw present.

3. Switch to external AC or battery power.

 

4. Change to another source of oxygen if

 

no other power source is available.

 

 

ACKNOWLEDGEMENT REQUIRED:

Press and hold the On/Off button to power down the unit and restart.

Invacare SOLO2

46

Part No 1156872

Page 46
Image 46
Invacare TPO100B user manual Make sure that car boat or motor