SECTION 5—TYPICAL PRODUCT PARAMETERS

MAXIMUM RECOMMENDED

3.0 LPM

 

FLOWRATE WITH 7KPA (1.01

 

 

PSI) BACKPRESSURE:

 

 

 

 

PRESSURE RELIEF

20 psi ± 5 psi (137.8 kPa ± 34.5 kPa)

ACTIVATION:

 

 

 

 

 

MAX OUTLET PRESSURE @

12.0 psig

 

3.0 LPM CONTINUOUS:

 

 

 

 

DIMENSIONS:

16.5 in high x 11 in wide x 8 in deep (41.9 cm

(WITHOUT CART)

high x 27.9 cm wide x 20.3 cm deep)

 

 

 

WEIGHT:

< 20 lbs (9.09 kg)

 

(UNIT WITHOUT CART)

 

 

 

 

 

BATTERY DURATION:

Pulse Mode

Continuous Mode

(TIMES ARE APPROXIMATE)

 

 

Setting 1 = 4.5 hrs

Setting 0.5 LPM = 4.5

 

 

 

hrs

 

 

 

 

Setting 2 = 3.5 hrs

Setting 1 LPM = 3.5 hrs

 

 

 

 

Setting 3 = 3.0 hrs

Setting 2 LPM = 2.5 hrs

 

 

 

 

Setting 4 = 2.5 hrs

Setting 3 LPM = 1.5 hr

 

 

 

 

Setting 5 = 2.5 hrs

 

 

 

 

BATTERY RECHARGE TIME:

5 hours

 

 

Recharge time increases if battery is charging

 

while unit is running.

 

 

 

 

HUMIDITY RANGE:

 

 

OPERATING HUMIDITY:

15% to 60% non condensing

TRANSPORT AND STORAGE:

Up to 95% non condensing

 

 

 

STANDARD TEMPERATURE

 

 

RANGE

 

 

(ALL POWER SOURCES)

 

 

OPERATING TEMPERATURE:

41°F to 95°F (5°C to 35°C)

TRANSPORT AND STORAGE

 

 

TEMPERATURE:

-2°F to 140°F (-20°C to 60°C)

 

 

 

EXTENDED TEMPERATURE

 

 

RANGE:

 

 

(USING AC OR DC POWER)

 

 

OPERATING TEMPERATURE:

95°F to 104° (35°C to 40°C)

AC SUPPLY:

Unlimited use all settings, all modes

DC SUPPLY:

Unlimited use all settings with pulse mode

 

Limited to 2.0 LPM, or less, with continuous

 

mode

 

 

 

 