SECTION 8—TROUBLESHOOTING GUIDE

 

 

 

LCD DISPLAY:

INDICATORS:

 

 

 

 

Continuous audible beep every half

 

second

 

RED indicator On.

 

GREEN indicator is Off.

 

 

 

DESCRIPTION:

SOLUTIONS:

 

 

 

 

1. With the unit turned Off, press each

 

control button looking for a stuck button.

 

2. Turn On the unit while ensuring that you

A stuck button is being detected upon

are not continuing to hold the Power

power up.

On/Off button once the indicator lights and

 

LCD begin to illuminate.

 

3. ONLY press the On/Off button during

 

startup.

 

 

 

ACKNOWLEDGEMENT REQUIRED:

 

 

 

 

 

Press and hold the On/Off button to power down the unit and restart. Call your

provider and report the problem if alarm continues.

 

 

 

 

 

 

LCD DISPLAY:

INDICATORS:

 

 

 

 

 

Triple audible beep every 10 sec

 

 

YELLOW indicator Flashing.

 

 

 

 

DESCRIPTION:

SOLUTIONS:

 

 

 

 

 

1. Connect transportable oxygen

 

Remaining battery capacity has fallen to

concentrator to either AC or DC power,

 

or use an alternate charged battery

 

15%. Battery requires charging.

 

module.

 

Battery gauge is empty.

 

2. If no other power source is available,

 

Loss of power is imminent.

 

Change to another source of oxygen if

 

 

alternate power is not available.

 

 

 

 

ACKNOWLEDGEMENT REQUIRED:

 

 

 

 

 

Press Return / Highlight button. Display will either show additional warnings / alarms or

 

revert to normal operating screens

 

 

 

 

 

Part No 1156872

41

Invacare SOLO2

Page 41
Image 41
Invacare TPO100B user manual Troubleshooting Guide