SECTION | ||
|
|
|
LCD DISPLAY: | INDICATORS: | |
|
|
|
| Continuous audible beep every half | |
| second | |
| RED indicator On. | |
| GREEN indicator is Off. | |
|
|
|
DESCRIPTION: | SOLUTIONS: | |
|
|
|
| 1. With the unit turned Off, press each | |
| control button looking for a stuck button. | |
| 2. Turn On the unit while ensuring that you | |
A stuck button is being detected upon | are not continuing to hold the Power | |
power up. | On/Off button once the indicator lights and | |
| LCD begin to illuminate. | |
| 3. ONLY press the On/Off button during | |
| startup. | |
|
|
|
ACKNOWLEDGEMENT REQUIRED: |
|
|
|
|
|
Press and hold the On/Off button to power down the unit and restart. Call your | ||
provider and report the problem if alarm continues. | ||
|
|
|
|
|
|
LCD DISPLAY: | INDICATORS: |
|
|
|
|
| Triple audible beep every 10 sec |
|
| YELLOW indicator Flashing. |
|
|
|
|
DESCRIPTION: | SOLUTIONS: |
|
|
|
|
| 1. Connect transportable oxygen |
|
Remaining battery capacity has fallen to | concentrator to either AC or DC power, |
|
or use an alternate charged battery |
| |
15%. Battery requires charging. |
| |
module. |
| |
Battery gauge is empty. |
| |
2. If no other power source is available, |
| |
Loss of power is imminent. |
| |
Change to another source of oxygen if |
| |
| alternate power is not available. |
|
|
|
|
ACKNOWLEDGEMENT REQUIRED: |
|
|
|
|
|
Press Return / Highlight button. Display will either show additional warnings / alarms or |
| |
revert to normal operating screens |
|
|
|
|
|
Part No 1156872 | 41 | Invacare SOLO2™ |