SECTION 8—TROUBLESHOOTING GUIDE

Warnings and Alarms

NOTE: In all cases, pressing and holding the On/Off button for 2‐3 seconds will turn off and reset the unit.

LCD DISPLAY:

INDICATORS:

 

 

 

Triple audible beep every 10 sec

 

YELLOW indicator Flashing.

 

 

DESCRIPTION:

SOLUTIONS:

 

 

(Only possible with Pulse Mode)

1. Verify the cannula is connected, not

Transportable oxygen concentrator has

kinked, properly positioned and you are

breathing through your nose.

not detected a breath for a predetermined

time period.

 

 

 

ACKNOWLEDGEMENT REQUIRED:

 

 

 

Press Return / Highlight button. Display will either show additional warnings / alarms or revert to normal operating screens. This warning will be turned off for a minimum of 5 minutes.

LCD DISPLAY:

INDICATORS:

 

 

 

Single audible beep every 15 sec

 

YELLOW indicator On.

 

 

DESCRIPTION:

SOLUTIONS:

 

 

(Only possible with Pulse Mode)

1. Immediately reduce your activity level to

Your breathing rate has begun to exceed

get your breath rate to slow down.

 

the capacity of the transportable oxygen

2. Temporarily switch to Continuous Flow

concentrator.

Mode.

 

 

ACKNOWLEDGEMENT REQUIRED:

 

 

 

A sufficient reduction of your breath rate will automatically turn off the warning.

Alternately, press Return / Highlight button. Display will either show additional warnings / alarms or revert to normal operating screens. This warning will be turned off for a minimum of 5 minutes.

Part No 1156872

37

Invacare SOLO2

Page 37
Image 37
Invacare TPO100B user manual LCD Display Indicators, Description Solutions, Acknowledgement Required