SECTION 8—TROUBLESHOOTING GUIDE

LCD DISPLAY:

INDICATORS:

 

 

 

Double audible beep every 15 sec

 

YELLOW indicator On.

 

 

DESCRIPTION:

SOLUTIONS:

 

 

Remaining battery capacity has fallen to

1. Connect transportable oxygen

concentrator to either the AC or DC

25%. Battery requires charging.

power, or use an alternate charged battery

Battery gauge is empty

module.

 

 

 

ACKNOWLEDGEMENT REQUIRED:

Press Return / Highlight button. Display will either show additional warnings / alarms or revert to normal operating screens.

LCD DISPLAY:

INDICATORS:

 

 

 

Single audible beep every 15 sec

 

YELLOW indicator On.

 

 

DESCRIPTION:

SOLUTIONS:

 

 

 

Unit is too hot, or too cold, to allow it to

1.

Move unit to warmer, or cooler,

surroundings. Allow unit to cool down to

operate.

less than 95° F (35° C), or warm up to 50°

 

Unit will run internal fans to help lower, or

F (10° C). Use AC or DC power.

raise, internal temperature.

2.

Clean intake filter.

 

Fans will turn off after 10 minutes

3.

Use back-up oxygen while waiting.

regardless of temperature.

4.

Turn unit Off, then On again to retry.

 

 

 

 

ACKNOWLEDGEMENT REQUIRED:

While the unit is warming / cooling the alarm screen will remain active. The unit will turn completely off (battery mode) or return to the standby screen (AC & DC mode), whenever the desired temperature is reached or 10 minutes of cooling has expired.