SECTION 8—TROUBLESHOOTING GUIDE

NO CHARGING SYMBOL SHOWN ON STANDBY SCREEN

SYMPTOM

PROBABLE CAUSE

SOLUTION

With the unit turned

Power source is no good, or

Try another power outlet

Off and AC or DC

there is a loose connection.

and check connections to

power plugged into

 

unit.

the unit, no charging

 

 

Battery is outside the allowed

Allow unit to cool down to

icon is displayed.

temperature range for charging.

less than 95° F

 

 

 

(35° C), or warm up to 50° F

 

 

(10° C).

 

 

 

 

Battery module is not fully

Ensure the battery module is

 

seated.

completely inserted.

 

 

 

 

Standby screen is not being

Press the On/Off button to

 

displayed.

display the Standby screen

 

 

when using DC power.

 

 

 

 

UNIT IS BEEPING WITH NO LIGHTS OR DISPLAY

 

 

 

SYMPTOM

PROBABLE CAUSE

SOLUTION

All power sources are

Battery pack has been removed

Install any battery pack

missing.

and there is no external power

available.

 

While operating without a

Beeping will last 10-15

 

battery, the external power is

minutes if no power source

 

lost.

is provided.

 

 

 

Invacare SOLO2

36

Part No 1156872

Page 36
Image 36
Invacare TPO100B user manual Unit is Beeping with no Lights or Display, Symptom Probable Cause Solution