SECTION
POWER INPUT: | 24 VDC @ 7.5 Amp |
| or |
| |
| (12.6 VDC Nom.) |
|
|
EXTERNAL POWER SUPPLY |
|
INPUT: |
|
AC POWER SUPPLY: | 120 VAC, 50/60 Hz @ 2.5 amps |
| 230 VAC, 50/60 Hz @ 1.25 amps |
DC POWER SUPPLY: | |
| (12.6 VDC Nom. @ 10.0 Amp max) |
|
|
SOUND LEVEL: | < 40 dBA weighted @ 2 LPM continuous and |
| all pulse settings |
|
|
ALTITUDE: | Up to 10,000 ft (3046 m) above sea level |
| Titration recommended for use above 10,000 |
| ft (3046 m) |
|
|
OXYGEN CONCENTRATION:* | 87% to 95.6%, after initial |
*BASED ON AN ATMOSPHERIC | (approximately 5 minutes) at all flow rates |
PRESSURE OF 14.7 PSI (101 KPA) AT |
|
70°F (21°C) NOMINAL |
|
|
|
CONSERVER TRIGGER | < 0.20 cmH2O max pressure drop |
SENSITIVITY: |
|
FACTORY SET - NO ADJUSTMENT | All settings - using 7 ft (2.1 m) cannula |
PRESSURE ACTIVATED |
|
|
|
CONSERVER TRIGGER DELAY: | 35 mSec max |
FACTORY SET - NO ADJUSTMENT |
|
| Nominal value - using 7 ft (2.1 m) cannula |
|
|
CONSERVER BOLUS | Setting 1: 400 cc |
DELIVERY: | Setting 2: 800 cc |
FIXED MINUTE VOLUME | Setting 3: 1200 cc |
| Setting 4: 1600 cc |
| Setting 5: 2000 cc |
| Total cc’s ± 75 cc delivered per minute |
|
|
CONSERVER BREATH RATE | Up to, and including, 35 BPM without |
CAPACITY: | reduction of bolus minute volume |
|
|
CONTINUOUS FLOW | 0.5 through 3.0 LPM @ 0 psi |
SETTINGS: | 0.5 LPM increments |
| All settings are ± 0.2 LPM |
| (2.0 LPM max w/Ext DC Power Cable) |
|
|
Part No 1156872 | 15 | Invacare SOLO2™ |