SECTION 8—TROUBLESHOOTING GUIDE

 

 

LCD DISPLAY:

INDICATORS:

 

 

 

Continuous audible beep every half

 

second.

 

RED indicator is On.

 

GREEN indicator is Off.

 

 

DESCRIPTION:

SOLUTIONS:

 

 

Abnormal compressor operation has been

1. Turn unit Off, then On again to retry.

detected.

2. Change to another source of oxygen if

alarm continues.

 

 

 

ACKNOWLEDGEMENT REQUIRED:

 

 

 

Press and hold the On/Off button to power down the unit and restart.

Call your provider and report the problem if alarm continues.

 

 

 

 

LCD DISPLAY:

INDICATORS:

 

 

 

Triple audible beep every 10 sec

 

YELLOW indicator Flashing.

 

 

DESCRIPTION:

SOLUTIONS:

 

 

Oxygen sensor feedback is out-of-range.

1. Turn unit Off, then On again to retry.

Sensor failure is likely

 

 

 

ACKNOWLEDGEMENT REQUIRED:

 

 

 

Press Return / Highlight button. Display will either show additional warnings / alarms or

revert to normal operating screens.

 

If the alarm is issued and acknowledged a second time, the audible alarm will be

deactivated until the unit is turned off.

 

Call your provider and report the problem.

 

 

 

Part No 1156872

45

Invacare SOLO2

Page 45
Image 45
Invacare TPO100B user manual Troubleshooting Guide