SECTION | ||
|
| |
LCD DISPLAY: | INDICATORS: | |
|
| |
| Continuous audible beep every half | |
| second. | |
| RED indicator is On. | |
| GREEN indicator is Off. | |
|
| |
DESCRIPTION: | SOLUTIONS: | |
|
| |
Abnormal compressor operation has been | 1. Turn unit Off, then On again to retry. | |
detected. | 2. Change to another source of oxygen if | |
alarm continues. | ||
| ||
|
| |
ACKNOWLEDGEMENT REQUIRED: |
| |
|
| |
Press and hold the On/Off button to power down the unit and restart. | ||
Call your provider and report the problem if alarm continues. | ||
|
| |
|
| |
LCD DISPLAY: | INDICATORS: | |
|
| |
| Triple audible beep every 10 sec | |
| YELLOW indicator Flashing. | |
|
| |
DESCRIPTION: | SOLUTIONS: | |
|
| |
Oxygen sensor feedback is | 1. Turn unit Off, then On again to retry. | |
Sensor failure is likely | ||
| ||
|
| |
ACKNOWLEDGEMENT REQUIRED: |
| |
|
| |
Press Return / Highlight button. Display will either show additional warnings / alarms or | ||
revert to normal operating screens. |
| |
If the alarm is issued and acknowledged a second time, the audible alarm will be | ||
deactivated until the unit is turned off. |
| |
Call your provider and report the problem. |
| |
|
|