SECTION
LCD DISPLAY: | INDICATORS: | |
|
| |
| Triple audible beep every 10 sec | |
| YELLOW indicator Flashing | |
|
| |
DESCRIPTION: | SOLUTIONS: | |
|
|
|
(Only possible with Continuous Flow | 1. | Verify the cannula is not kinked. |
Mode) | 2. | Replace outlet filter, if necessary. |
Output flow is more than 0.3 LPM below | 3. | Turn unit Off. Wait one minute, then |
the flow setting. | retry. | |
|
|
|
ACKNOWLEDGEMENT REQUIRED:
Press Return / Highlight button. Display will either show additional warnings / alarms or revert to normal operating screens. This warning will be turned off for a minimum of 15 minutes.
If the warning is issued and acknowledged a second time, the warning will be deactivated until the unit is turned off.
Call your provider and report the problem if the warning continues with next use.
LCD DISPLAY: | INDICATORS: | |
|
| |
| Single audible beep every 15 sec | |
| YELLOW indicator On. | |
|
| |
DESCRIPTION: | SOLUTIONS: | |
|
| |
| 1. Verify unit is in recommended | |
| environmental temperature. Refer to | |
The oxygen output purity has fallen to a | Typical Product Parameters on page 14. | |
2. Clean intake filter and ensure both | ||
value between 73% and 85% (± 1%). | ||
| intake and exhaust are not blocked | |
| 3. Turn unit Off, then On again to retry. | |
|
|
ACKNOWLEDGEMENT REQUIRED:
Press Return / Highlight button. Display will either show additional warnings / alarms or revert to normal operating screens.
If the warning is issued and acknowledged a second time, the warning will be deactivated until the unit is turned off.
Call your provider and report the problem if the warning continues with next use.
Part No 1156872 | 39 | Invacare SOLO2™ |