SECTION 8—TROUBLESHOOTING GUIDE

LCD DISPLAY:

INDICATORS:

 

 

 

Ten audible beeps every 10 sec

 

RED indicator Flashing Fast.

 

 

DESCRIPTION:

SOLUTIONS:

 

 

 

1. Connect transportable oxygen

Remaining battery capacity has been

concentrator to either AC or DC power,

or use an alternate charged battery

completely used. Battery requires charging.

Unit is turning off!

module.

2. If no other power source is available,

 

 

Change to another source of oxygen.

 

 

ACKNOWLEDGEMENT REQUIRED:

None required. Unit turns off within 30 seconds. Find alternate power and restart normally.

LCD DISPLAY:

INDICATORS:

 

 

 

Triple audible beep every 10 sec

 

YELLOW indicator Flashing.

 

 

DESCRIPTION:

SOLUTIONS:

 

 

 

 

1.

Move unit to warmer, or cooler,

Unit is too hot, or too cold, to allow it to

surroundings. Allow unit to cool down to

less than 95° F (35° C), or warm up to 50°

continue to operate.

F (10° C). Use AC, or DC, adapter for

 

Unit will continue to run internal fans to

power.

help lower, or raise, internal temperature.

2.

Clean intake filter.

Fans and unit will turn off after 10 minutes

3.

Turn unit Off, then On again to retry.

regardless of temperature.

4.

Change to another source of oxygen

 

 

while waiting

 

 

 

ACKNOWLEDGEMENT REQUIRED:

While the unit is warming / cooling the alarm screen will remain active. The unit will turn completely off (battery mode) or return to standby screen (AC & DC mode), whenever the desired temperature is reached or 10 minutes of cooling has expired.