SECTION 6—OPERATING INSTRUCTIONS
Invacare SOLO2 26 Part No 1156872
Breath Detect Indication
(PulseModeOnly)TheGREENindicatorlightwillblinkoffeachtimean
inhalationisdetectedandthetransportableoxygenconcentratoroutputsa
measuredpulseofoxygen.
Audible Alarm
Thetransportableoxygenconcentratorisequippedwithanaudiblealarm.
TheaudiblealarmwillactivatedinconjunctionwithIndicatorsandLCD
DisplaytoannunciateAlarmsandWarni ngs.
Thetransportableoxygenconcentratorwillalsoissueasteadybeepwhen
allpowersourcesareremoved.Thebeepingwillstopafter1020minutesor
onceapowersourceisconnected.
Thetransportableoxygenconcentratorwillbrieflyturnonallthreelights
andtheLCDdisplayandcheckforstuckbuttonsuponpowerup.Itwill
thenoutputabriefaudiblebeeptoindicatetheunitisoperational.
Control/Adjustment Buttons
TherearefourbuttonslocatedsurroundingtheLCDdisplay.Theyareused
toturntheuniton/off,toproceedtothenextscreen,toselectandhighlight
adjustablesettingsandtoadjustthosesettings.
On/Off Button
Toturnthetransportableoxygenconcentratoronoroff,pressandholdthe
On/Offbuttonforaperiodof23seconds,nottoexceed5seconds.Releasing
thebuttontoosoonwillnotallowtheunittodoeither.Ifthebuttonisheld
toolongwhenturningontheunit,theStuckButtonalarmcouldbe
activated.
Mode/Return/Highlight Button
Pressingbuttonfor12secondsintheStandbyscreen(FIGURE6.5on
page27)willswitchtoNormalOperatingscreen(FIGURE6.6onpage 27).
Pressingthisbuttonfor1secondintheNormalOperatingscreenwilltoggle
betweenthecontinuousandpulsemodes.
Pressingthisbuttonfor23secondsintheNormalOperatingscreenwill
displaytheLCDAdjustmentscreen.IntheLCDAdjustmentscreen
pressingthisbuttonfor1secondwillswitchcolumnsonthescreen.Tosave
changesandexittheLCDadjustmentscreenthehighlightbarneedstobe
positionedintheleftcolumn,thenpressthisbuttonfor23secondstoreturn
totheNormalOperatingscreen.Ifthescreenswitcheswithoutpressingand
holdingthebuttonfor23secondsthechangesmadeintheLCD
Adjustmentscreenwillnotbesaved.
Increase and Decrease Buttons
Whenthehighlightbarisonthesettingname,thesebuttonsareusedto
selectthenextsettinginthelist.Whenthehighlightbarisonthesetting
value,thesebuttonswillchangethevaluetothenextvalue available.