Image calibration | Use the calibration target provided with your scanner. Be sure to use a | |
| good, clean calibration target. Additional calibration targets can be | |
| ordered. See Chapter 5, “Supplies and accessories” for ordering | |
| information. | |
| NOTE: The screens shown in this section are for the TWAIN | |
|
| Datasource. The screens displayed on your system may be |
|
| different. |
| 1. | Open the pod and clean the imaging guides. Refer to the |
|
| maintenance procedures in Chapter 5. |
| 2. | If the lamps have not been on, allow the lamps to warm up for 90 |
|
| seconds. |
| 3. | Center the side guides in the elevator tray and output tray. |
| 4. | Place the calibration target in the elevator tray. |
| 5. | Open the Scan Validation Tool. |
6.Select Kodak Scanner: i600 or Kodak Scanner: i700. The main Kodak Scanner window will be displayed.