Image calibration

Use the calibration target provided with your scanner. Be sure to use a

 

good, clean calibration target. Additional calibration targets can be

 

ordered. See Chapter 5, “Supplies and accessories” for ordering

 

information.

 

NOTE: The screens shown in this section are for the TWAIN

 

 

Datasource. The screens displayed on your system may be

 

 

different.

 

1.

Open the pod and clean the imaging guides. Refer to the

 

 

maintenance procedures in Chapter 5.

 

2.

If the lamps have not been on, allow the lamps to warm up for 90

 

 

seconds.

 

3.

Center the side guides in the elevator tray and output tray.

 

4.

Place the calibration target in the elevator tray.

 

5.

Open the Scan Validation Tool.

6.Select Kodak Scanner: i600 or Kodak Scanner: i700. The main Kodak Scanner window will be displayed.

3-12

A-61500 May 2010

Page 28
Image 28
Kodak 6J7611A manual Image calibration