Field Calibration
The field calibration on this unit is done using the Field Calibration System.
Periodic field calibration is recommended to maintain measurement precision.
Note
•Always carry out field calibration when changing lenses.•The Field Calibration System that corresponds to the type of lens the unit is equipped with must be used when con- ducting field calibration.
•This unit and the Field Calibration System are paired, so it is not possible to correctly calibrate it when the serial numbers differ. Use the unit and Field Calibration System combination that have the same serial number.
Memo
The type of lens the unit is equipped with and the serial number of the unit can be confirmed in the Status information.
Ref. For details on displaying the status information, refer to page 41.
Field Calibration System
The Field Calibration System is comprised of the Unit Section which contains the Field Calibration System, which corresponds to the type of lens on the unit, and the Frame Section, which maintains a facing position at a set distance from the unit.
• Parts Nomenclature
Lock Lever B
Lock Lever A
Handle B | Handle A |
| Unit Section (Field Calibration System) |
| * Equipped with field calibration systems for the |
| WIDE lens type and MIDDLE/TELE lens type. |
Unit Section Cover
Unit Attachment Section
Frame Section
(Frame for Calibration System)
Support
Section
Main Unit Attachment
Section
Rubber Foot (⋅6): With Height Adjustment Mechanism
*The rubber feet adjustment mechanism has a structure shown in the drawing below.
(Frame section)
Stopper
Lock nut
Rubber foot block [Specifications]
•Lowest height: 43mm
•Adjustable range: Max. 10mm (with no preventive structure)
32