Error Messages

The following error messages appear if incorrect operation is performed or an abnormality occurs with the VIVID 9i. If an error message appears, take the appropriate corrective actions as shown in the table below. If the problem still does not disappear, contact a KONICA MINOLTA SENSING authorized service facility.

Error Message

Meaning

Action
(Alphabetically)

 

 

AF ERROR

Auto-focus failed due to low contrast

Set the manual focus.

LOW CONTRAST

of the object. Press any key.

 

PRESS ANY KEY

 

 

BACKUP ERROR

Failed to back up the data.

All the settings will be initialized, so set

PRESS ANY KEY

Press any key.

them again.

BACKUP ERROR

Failed to back up field calibration

Field calibration values have been reini-

PRESS ANY KEY AND SCAN

values.

tialized. Please try field calibration again.

FIELD CALIBRATION SYS-

Press any key and scan field calibra-

 

TEM

tion system.

 

CAUTION

The laser beam is not emitted prop-

Highly dangerous! Turn off the power

LASER SCANNER IS

erly. Turn off the power immediately.

immediately and contact a KONICA

OUT OF ORDER

 

MINOLTA SENSING authorized service

PLEASE TURN OFF

 

facility.

CHANGE LENS

The lens is now being replaced.

Attach the lens hood properly.

S/N ********

Press any key and scan field calibra-

 

Check the serial No. of the lens, and

PRESS ANY KEY AND SCAN

tion system.

replace the lens properly.

FIELD CALIBRATION SYS-

 

 

TEM

 

 

LIGHT IMPROPER

Inadequacy amount of light

Illuminate the object brighter and scan it.

SAME FILE NAME

A file of the same file name already

Specify another file name.

 

exists.

 

SYSTEM ERROR

A system error has occurred.

Turn OFF the power, then turn it ON

PLEASE TURN OFF

Turn OFF the power.

again. If the problem still does not disap-

 

 

pear, contact a KONICA MINOLTA

 

 

SENSING-authorized service facility.

TERMINATE ERROR

SCSI terminator error has occurred.

Set the terminator to “ENABLE”.

PRESS ANY KEY

Press any key and set the terminator

 

SET TERMINATE ENABLE

to “ENABLE”.

 

WB ERROR

Failed to adjust white balance.

Adjust white balance in a bright place.

LIGHT IMPROPER

Press any key.

(White balance in effect before occur-

PRESS ANY KEY

 

rence of this error has not been updated.)

48