Field Calibration
6 Check that field calibration was successful.If you are using field calibration results to measure the dimensions of the field calibration system, the software shows the differences from the measurements made at time of shipping. Differences are shown for four places: X1, Y1, X2, and Y2.
•If field calibration was not successful, one of the following messages appears. Proceed in accordance with the displayed message.
Message | Meaning |
|
| What To Do | ||
Field Calibration: | Failed | to | recognize | • | Remove the laser barrier and the receiving lens cap. | |
Scan Error | the field | calibration | • | Check that you are using the correct system type for the lens | ||
| system. |
|
|
|
| used. |
|
|
|
|
| • | Check that foreign matter is not in the “Calibration Area”. |
|
|
|
|
| • | Check that there is no effect of light from outside. |
|
|
|
|
| • Be sure that the Field Calibration Chart unit is correctly | |
|
|
|
|
|
| mounted to the Frame section. |
|
|
|
|
| • | Check that the image of the Field Calibration Chart is cor- |
|
|
|
|
|
| rectly displayed in the view angle of the finger of the VIVID |
|
|
|
|
|
| |
|
|
|
|
|
| gon Editing Tool. |
Field Calibration: | Failed | to | correctly | • | Check that you are using the correct system type for the lens | |
Evaluate Error | measure | the | field |
| used. | |
| calibration system. | • | Check that foreign matter is not in the “Calibration Area”. | |||
|
|
|
|
| • | Check that there is no effect of light from outside. |
|
|
|
|
| • Be sure that the Field Calibration Chart unit is correctly | |
|
|
|
|
|
| mounted to the Frame section. |
|
|
|
|
| • | Check that the image of the Field Calibration Chart is cor- |
|
|
|
|
|
| rectly displayed in the view angle of the finger of the VIVID |
|
|
|
|
|
| |
|
|
|
|
|
| gon Editing Tool. |
Field Calibration: | Adequate | field | cali- | • | Remove the lens, and then set it again correctly. | |
Calibrate Error | bration | results | can- | • | Check that foreign matter is not in the “Calibration Area”. | |
| not be obtained. |
| • | Check that there is no effect of light from outside. | ||
|
|
|
|
| • Be sure that the Field Calibration Chart unit is correctly | |
|
|
|
|
|
| mounted to the Frame section. |
|
|
|
|
| • | Check that the image of the Field Calibration Chart is cor- |
|
|
|
|
|
| rectly displayed in the view angle of the finger of the VIVID |
|
|
|
|
|
| |
|
|
|
|
|
| gon Editing Tool. |
|
|
|
|
| • When multiple units of the instrument are used, check to | |
|
|
|
|
|
| make sure that the Field Calibration System matches the |
|
|
|
|
|
| serial number of the instrument. |
|
|
|
|
| • | If there is any dirt on the laser window or the receiving lens, |
|
|
|
|
|
| clean these as explained on page 3. |
|
|
|
|
| • | Clean off any dust on the field calibration system. |
|
|
|
|
| • | If the field calibration system is scratched or dented, contact |
|
|
|
|
|
| a KONICA MINOLTA SENSING authorized service facil- |
|
|
|
|
|
| ity. |
|
|
|
|
| • | If the same error recurs when you retry the field calibration, |
|
|
|
|
|
| contact a KONICA MINOLTA SENSING authorized ser- |
|
|
|
|
|
| vice facility. |
7 | Click the [Cancel] button. |
| The dialog closes, and field calibration is completed. |
39