Field Calibration
• Installation Space
A secure installation location that meets the dimensional requirements as shown in the drawing below is necessary to install the Field Calibration System. Before using the Field Calibration System, a “Calibration Area” for stable and high quality calibration is required, in addition to the place that is necessary to position the Field Calibration System.
Approx. 73 cm |
|
|
|
Installation Space | |||
|
|
|
|
Approx. 145 cm
Approx. 46 cm
• Calibration Area
The “Calibration Area” is the box shaped area shown in the drawing below. The dimensions are:
•30 cm or less in front of or behind the Chart unit
•20 cm or less right or left to the Chart unit
•20 cm or less above the Chart unit
Note
•Prevent foreign matter from getting into the “Calibration Area”. When executing Field Calibration, do not allow workers into the area. It may result in incorrect calibration.
•Be careful to block incident light from the outside. Execute Field Calibration under fluorescent light at less than 500lx.
•Prevent light from windows.
Calibration Area
In 20 cm above the Chart unit
In 30 cm front of
or behind the Chart unit
In 20 cm right or left to the Chart unit
33