5.Working with the Instrument

Key/setting dial

semiautomatic sectioning mode

automatic sectioning mode

 

 

 

 

The blade feed rate can be adjusted

ditto

1

from 0.01 – 1.5 mm/s:

 

 

 

 

0.01 – 0.1 in 0.01 mmm/s increments,

 

 

0.10 – 0.5 in 0.02 mmm/s increments,

 

 

0.50 – 1.5 in 0.10 mmm/s increments.

 

1

Adjusting the amplitude from

ditto

0 – 3 mm in 0.05 mm increments

 

 

 

1

Not possible.

Setting the section thickness for

 

 

automatic mode – max. 1000μm.

Display for the current specimen ditto holder position (lowest position =

0μm, uppermost position = 20,000μm.) You can reset the display to zero at any point using the CLEAR/μm key. The section thicknesses are then added in the μm display.

24

Operating Manual V1.1 – 05/2007

Page 24