5. Working with the Instrument

5.2.4 Selecting the Sectioning Parameters

SPEED: Blade feed rate – upper display – LED mm/s. The desired blade feed rate can be adjusted from 0.01 – 1.5 mm/s using rotary knob 1:

0.01 – 0.1 in 0.01mm/s increments

0.10 – 0.5 in 0.02mm/s increments

0.50 – 1.5 in 0.10mm/s increments

AMPL: upper display – LED mm: displays the amplitude in mm: from 0 - 3 mm in 0.05mm increments

Starts the sectioning process

Starts the sectioning process using the selected amplitude and blade feed rate. You can start the sectioning process by pressing the RUN/STOP key a second time, or you can stop the process immediately by pressing the "Blade back" or "Blade forward" key. The μm display is then reset to "0".

To start a new sectioning process, use the "Blade back" key to move the blade to the beginning of the specimen, set the desired section thickness and restart the sectioning process.