5.Working with the Instrument
5.3Control Element on the Control Panel of the VT1200S
The Leica VT1200S is a fully automatic microtome with a vibrating blade that can be operated in semiautomatic or automatic sectioning mode.
Key/setting dial | semiautomatic sectioning mode | automatic sectioning mode |
|
|
|
Switch on the instrument.
1 = on 0 = off
When the instrument is switched | ditto | ||
on, to make it easier to insert a |
|
| |
specimen, the specimen recepta- |
|
| |
cle automatically moves into the |
|
| |
lowest position (the LED in the |
|
| |
DOWN key lights up) and the blade |
|
| |
holder moves into the rearmost po- |
|
| |
sition (the LED in "Blade away" from |
|
| |
the specimen lights up). |
|
| |
If semiautomatic sectioning mode |
|
| |
has been selected before the unit |
|
| |
is switched off, the following saved |
|
| |
parameters are called up again |
|
| |
once the unit is switched back on: | • | Blade feed rate (SPEED) | |
• | Blade feed rate (SPEED) | • | Selected amplitude (AMPL) |
• | Selected amplitude (AMPL) | • | Saved section thickness |
|
|
| (AUTO FEED) |
The LED illumination is switched on | ditto |
automatically. You can switch it off |
|
using the ON/OFF key. |
|
The LED MAN is active. semiautomatic sectioning mode is enabled.
The LED AUTO is active.
The automatic sectioning mode is enabled.
22 | Operating Manual V1.1 – 05/2007 |