5.Working with the Instrument
Key/setting dial | semiautomatic sectioning mode | automatic sectioning mode |
|
|
|
| ➞ Rotary knob 2 |
1 | Mode: Choice between AUTO and |
| MAN with rotary knob 1; for |
| semiautomatic sectioning mode, |
2 | MAN must be selected. |
|
➞Rotary knob 2
Mode: Choice between AUTO and MAN with rotary knob 1; for the automatic sectioning mode, AUTO must be selected.
| ➞ Rotary knob 2 |
|
| ➞ Rotary knob 2 | |
1 | Stroke type (CUT) – only single | Stroke type (CUT): Choice between | |||
| stroke (SINGLE) can be selected | single stroke (SINGLE) and contin- | |||
| with rotary knob 1. If continuous | uous stroke (CONT) with rotary | |||
2 | stroke (CONT) is selected, an audi- | knob 1. | |||
| ble warning signal sounds. |
| |||
|
|
|
|
| |
| ➞ Rotary knob 2 |
|
| ➞ Rotary knob 2 | |
1 | Specimen retraction (RETRACT) | The specimen retraction (RETRACT) | |||
| cannot be set. |
|
| can be set from 0 – 100μm in 10μm | |
|
|
|
|
| increments with rotary knob 1. |
2 |
| Value | cannot be |
|
|
| changed. NO automat- |
|
| ||
|
|
|
| ||
|
| ic retraction is possi- |
|
| |
|
| ble in | semiautomatic |
|
|
|
| sectioning mode. |
|
| |
|
|
|
|
|
|
|
|
|
|
| |
1 | ➞ Rotary knob 2 |
|
| ditto | |
The LED illumination can be adjust- |
| ||||
|
|
ed to 5 different levels of brightness using rotary knob 1.
2
30 | Operating Manual V1.1 – 05/2007 |