7. Malfunctions: Meanings and Troubleshooting

Error/INF No.

Error

Troubleshooting

Remark

Error 27

- "Bottom" limit switch

- Check whether an obsta-

 

 

of the Z-axis is not

cle is blocking the travel

 

 

reached during initial-

path of the ice bath.

 

 

ization or operation.

Remove the obstacle.

 

 

 

If there is no obstacle,

 

 

 

and the error message

 

 

 

persists after the in-

 

 

 

strument is switched on

 

 

 

again: inform service.

 

Error 28

- The "top" limit switch

 

of the Z-axis is not

 

reached.

- Check whether an obsta- cle is blocking the travel path of the ice bath.

Remove the obstacle. If there is no obstacle, and the error message persists after the in- strument is switched on again: inform service.

Error 31

- Both X sensors activat-

-

Inform service!

 

ed (during initialization

 

 

 

or in normal operation)

 

 

 

 

 

 

Error 32

- Both Z sensors activat-

-

Inform service!

 

ed (during initialization

 

 

 

or in normal operation)

 

 

After all error messages, the instrument MUST be switched off, then on again using the main switch.