7.Malfunctions: Meanings and Troubleshooting

Error/INF No.

Error

Troubleshooting

Remark

InF 41

- Control board (C1/back

 

panel) and keyboard

 

(C2) have different

 

software versions.

-It is possible that due to different software ver- sions, some or all func- tions of the instrument may be partly or entirely unavailable.

Inform service and have the software updated to the latest version.

InF 42

- Control board (C1/back

 

panel) and VibroCheck

 

(C3) have different

 

software versions.

-It is possible that due to different software ver- sions, some or all func- tions of the instrument may be partly or entirely unavailable.

"InF" messages can be removed by pressing the Clear key.

The message appears once each time the instrument is switched ON. "InF" messages do not cause the instrument to be locked.

After all error messages, the instrument MUST be switched off, then on again using the main switch. – Exceptions are InF 41 and 42.

Error 51

- Horizontal calibration

- Possible operating error,

 

of VibroCheck is not

see Operating Manual.

 

possible.

- Great deal of damage or

 

 

dirt on the blade or blade

 

 

holder. Use a new blade

 

 

or blade holder.

 

 

- Controllers have different

 

 

software versions. Have

 

 

the software updated.

 

 

- VibroCheck is defective.

 

 

Have it checked by ser-

 

 

vice.

- A calibration is car- ried out even if the blade is jagged.

A great deal of obvi- ous damage or dirt must be present before a blade cali- bration is canceled automatically.

Leica VT1200/VT1200S – Microtome

43

Page 43
Image 43