7.Malfunctions: Meanings and Troubleshooting
Error/INF No. | Error | Troubleshooting | Remark |
InF 41 | - Control board (C1/back |
| panel) and keyboard |
| (C2) have different |
| software versions. |
-It is possible that due to different software ver- sions, some or all func- tions of the instrument may be partly or entirely unavailable.
Inform service and have the software updated to the latest version.
InF 42 | - Control board (C1/back |
| panel) and VibroCheck |
| (C3) have different |
| software versions. |
-It is possible that due to different software ver- sions, some or all func- tions of the instrument may be partly or entirely unavailable.
"InF" messages can be removed by pressing the Clear key.
The message appears once each time the instrument is switched ON. "InF" messages do not cause the instrument to be locked.
After all error messages, the instrument MUST be switched off, then on again using the main switch. – Exceptions are InF 41 and 42.
Error 51 | - Horizontal calibration | - Possible operating error, |
| of VibroCheck is not | see Operating Manual. |
| possible. | - Great deal of damage or |
|
| dirt on the blade or blade |
|
| holder. Use a new blade |
|
| or blade holder. |
|
| - Controllers have different |
|
| software versions. Have |
|
| the software updated. |
|
| - VibroCheck is defective. |
|
| Have it checked by ser- |
|
| vice. |
- A calibration is car- ried out even if the blade is jagged.
A great deal of obvi- ous damage or dirt must be present before a blade cali- bration is canceled automatically.
Leica VT1200/VT1200S – Microtome | 43 |