5.

Working with the Instrument

 

 

 

Key/setting dial

semiautomatic sectioning mode

automatic sectioning mode

 

 

 

 

Rotary knob 2

ditto

1

The feed rate (FOR/REV) for the

 

 

"Blade forward" and the "Blade

 

 

back" keys can be set between 1

 

2

and 5 mm/s in 0.5 mm/s increments

 

 

 

 

with rotary knob 1.

 

 

 

 

1

Rotary knob 2

ditto

Switches the vibration movement

 

 

 

 

(FOR/VIB) of the "Blade forward"

 

 

key either on or off using rotary

 

2

knob 1.

 

 

 

 

 

 

ditto

1

Rotary knob 2

 

 

Key acknowledgment (BEEP) on or

 

 

off with rotary knob 1.

 

2

 

 

 

 

 

 

Save the parameter and leave by

ditto

 

pressing the menu MENU key.

 

Leaving the menu and saving the parameter is possible at any point in the menu.

If you want to call up the parameters of a certain user (for example, user 3), press the Menu key and select user 3. Then confirm by pressing the Menu key 2x. The parameters stored under user3 are now enabled.