5.Working with the Instrument

Key/setting dial

semiautomatic sectioning mode

automatic sectioning mode

 

 

 

 

If the DOWN key is pressed a sec-

ditto

 

ond time while the specimen

 

 

receptacle is moving downwards,

 

 

the specimen receptacle comes to

 

 

a stop and the current position ap-

 

 

pears on the μm display (lowest

 

 

position = 0, uppermost position =

 

 

20,000μm). The μm display does

 

 

not change while the specimen

 

 

receptacle is in motion.

 

 

 

 

 

When the UP key is pressed and

ditto

 

held down, the specimen recepta-

 

 

cle moves quickly upwards into the

 

 

desired position. After the UP key

 

 

has been released, the current po-

 

 

sition of the specimen receptacle

 

 

appears on the μm display. If the

 

 

specimen receptacle reaches the

 

 

upper end position, the LED in the

 

 

UP key lights up (uppermost posi-

 

 

tion = 20,000μm). The μm display

 

 

does not change while the speci-

 

 

men receptacle is in motion.

 

Blade forward

Blade back

The "Blade forward" and "Blade

ditto

back" keys have to be held down

 

until the desired position is

 

reached. The blade feed rate can

 

be set in the menu: 1 – 5 mm/s in

 

0.5mm/s increments. Each time an end point is reached, the corre- sponding LED lights up in the key.

26

Operating Manual V1.1 – 05/2007

Page 26
Image 26
Leica Leica VT1200S manual Blade forward and Blade