7. Malfunctions: Meanings and Troubleshooting

Error/INF No.

Error

Troubleshooting

Remark

Error 52

-

Basic calibration of

-

Emitter diode or receiver

 

 

 

light barrier is not pos-

 

is defective.

 

 

 

sible for VibroCheck.

-

VibroCheck is defective.

 

 

 

 

-

Inform service!

 

 

 

 

 

 

 

Error 53

-

VibroCheck blade

-

No blade and/or blade

- VT1200 drives to the

 

 

search not successful.

 

holder swung out or not

lowest Z position us-

 

 

 

 

installed.

ing the stepper motor

 

 

 

-

Emitter diode or receiver

after an unsuccessful

 

 

 

 

dirty. Clean

search. (Allows blade

 

 

 

-

VibroCheck is defective.

installation)

 

 

 

 

 

 

Communication

-

Communication error

-

VT1200S: display lights up,

 

between keyboard

 

between control unit

 

but remains blank.

 

– VT1200/S

 

and VT.

-

VT1200: row of dots

 

 

 

 

 

flashes.

 

 

 

 

-

Inform service!

 

 

 

 

 

 

 

- The knife/blade

- The buffer tray was not

collides with the

pushed in up to the

buffer tray

rear pin or was not

 

clamped down during

 

installation.

Push in the buffer tray as far as it will go and clamp it down with the lever.

- The VibroCheck

- The VibroCheck was

collides with the

not pushed all the way

blade

in or was not clamped

 

down during installa-

 

tion.

-Push in the VibroCheck as far as it will go and clamp it down with the lever.

-Switching from SINGL - In MAN mode, only single cuts are possible. to CONT is not possible.

- The cutting window

-

In MAN mode, only single cuts are possible.

cannot be defined.

-

If multiple cuts are required, switch to AUTO mode.