5.5Adjustment of the Circular Level

On the instrument step-by-step

TPS12_030

Step Description

1.Level up the instrument in advance with the electronic level, assuming that the electronic level is correctly adjusted. SHIFT F12 to access STATUS Level & Laser Plummet.

2.The bubble must be centered. If it extends beyond the circle, use the allen keys supplied to centre it with the adjustment screws. Turn the instrument slowly 200 gon (180°). Repeat the adjustment procedure if the bubble does not stay centered.