Leica TPS1200 user manual Tools Adjustment Accuracy

Models: TPS1200

1 226
Download 226 pages 15.99 Kb
Page 98
Image 98

Check & Adjust

TPS1200

98

Step Description

5.MEAS (F1) to measure and to continue to the next screen.

Motorised instruments change auto- matically to the other face.

180°

Non-motorised instruments guide to

 

the other face.

 

The fine pointing has to be

180°

performed)

manually in both faces.

TPS12_025

6.TOOLS Combined II

MEAS (F1) to measure the same target in the other face and to calculate the instrument errors.

)If one or more errors are bigger than the predefined limits, the procedure has to be repeated. All measurements of the current run are rejected and none of them is averaged with the results from previous runs.

7.TOOLS Adjustment Accuracy

Page 98
Image 98
Leica TPS1200 user manual Tools Adjustment Accuracy