Step

Description

 

<No.of Meas:> Shows the number of runs executed. One run consists of

 

a measurement in face I and face II.

 

<σ l Comp:> and similar lines show the standard deviations of the deter-

 

mined adjustment errors. The standard deviations can be calculated from

 

the second run onwards.

 

 

)

It is recommended to measure at least two runs.

8.

MEAS (F5) if more runs have to be added. Continue with step 3.

 

OR

 

CONT (F1) to accept the measurements and to proceed to TOOLS

 

Adjustment Results. No more runs can be added later.

 

 

 

 

Check & Adjust

TPS1200

99

Page 99
Image 99
Leica TPS1200 user manual Second run onwards