The control solution test follows the normal test procedure, using control solution instead of blood.

Step 1: Turn on the meter.

The code number, time, and date will appear. The code number should match the code number on the test strip vial.

CODE 10

7:50AM 1/06/00

M

C

Step 2: Insert a test strip.

When INSERT CODE 10 (for exam- ple) STRIP appears, slide a test strip into the test strip holder, notched end first with the test spot facing up. Make sure you push the test strip in all the way until it stops.

INSERT

M

C

18