MESSAGE

PROBLEM

WHAT TO DO

ERROR 1

The sample was applied before

Repeat the test with a new test strip.

RETEST

the words APPLY SAMPLE

 

 

 

appeared.

 

 

 

 

 

 

ERROR 2

1. The test strip moved during

1.

Repeat the test with a new

RETEST

the test.

test strip.

 

2. The test strip was not insert-

2.

Repeat the test with a new

 

ed correctly.

test strip.

 

3. The test strip was removed

3.

Repeat the test with a new test

 

before the test was completed.

strip.

 

4. There was not enough blood

4.

Repeat the test with a new

 

on the test strip.

test strip.

 

5. The meter was used in very

5.

Move the meter away from the

 

bright light.

light source. Repeat the test with a

 

 

new test strip.

58

Page 59
Image 59
Lifescan Blood Glucose Monitor manual Error