MESSAGE

PROBLEM

WHAT TO DO

CLEAN

1. There is dirt, blood, or lint

1. Clean the test area according to

TEST

on the test area.

instructions.

AREA

2. Your hand or an object cov-

2. Repeat the test. Keep test area

 

 

ered the test area during testing.

clear.

 

3. The meter was used in very

3. Move the meter away from the

 

bright light.

light source. Repeat the test with a

 

 

new test strip.

 

 

 

REMOVE

The test strip was inserted

Remove the strip. Wait for the word

STRIP

before the word INSERT

INSERT before inserting test strip.

 

appeared on the display.

 

 

 

 

100 MG/DL or

If the word CONTRL appears

Repeat the test with a new test strip

5.6 MMOL/L

after a blood test, your blood

and apply a larger drop of blood.

(for example)

sample was too small, smeared,

 

CONTRL

or another drop was added after

 

8:01AM 6-01-00

the test began.

 

57